Panelized Walk-in Temperature & Humidity Test Chamber
طلب عرض سعر
الموقع الإلكتروني: http://testchamber-sa.com/6-1-panelized-walk-in-test-chamber.html
Environmental chamber designed with insulated panels for easy assembly and installation, and used for testing various products
Mold growth testing function
Application: Enables mold growth and cultivation tests to simulate long-term exposure to humid environments.
Infrared radiation simulation
Application: Enables testing of how samples withstand prolonged infrared radiation, helping assess material durability under thermal stress.
Features
- Customizable in size, temperature range, and ramp speed, with internal volumes available up to 2,000 m³
- Equipped with an electronic expansion valve to optimize cooling performance and reduce energy use
- Optional configurations tailored for industries such as photovoltaics, automotive, semiconductors, and new energy systems
Technical Parameters
| Model | | | | | | | | | | | | ||
| Chamber volume (m³) | 3.9/3.4 | 6.3/5.5 | 7.8/6.8 | 11.7/10.2 | 15.6/13.6 | 17.5/15.3 | 23.3/20.4 | 29.2/25.5 | 31.1/27.2 | 38.9/34.0 | 46.7/40.8 | ||
| Performance | Temperature range | -70℃~ 85℃ (customizable) | |||||||||||
| Humidity range (only models with humidity control) | (20~98)%RH/(20~75)℃ | ||||||||||||
| Temperature fluctuation | 0.6℃ | ||||||||||||
| Temperature deviation | ±2.0℃ | ||||||||||||
| Humidity deviation (only models with humidity control) | ±3.0%RH(>75%RH) ±5.0%RH(≤75%RH) | ||||||||||||
| Temperature rise rate | 25℃→ 85℃ | ||||||||||||
| ≤25min | ≤28min | ≤28min | ≤30min | ≤31min | ≤32min | ≤34min | ≤36min | ≤38min | ≤41min | ≤45min | |||
| Temperature drop rate | 25℃→rated extremely low temperature | ||||||||||||
| 30min-180min (depending on different models and configurations) | |||||||||||||
| Internal dimensions (mm) | W | 1800 | 2700 | 3600 | 2700 | 3600 | |||||||
| H | 2100 | ||||||||||||
| D | 900 | 1450 | 1800 | 1800 | 1800 | 2700 | 3600 | 4500 | 3600 | 4500 | 5400 | ||
Performance specifications are based on an ambient temperature between 15 °C and 35 °C, relative humidity up to 85%, and cooling water temperature not exceeding 28 °C. Measurements were taken with the chamber empty.
Certain models are compatible with standards such as PV1200/PV2005, PR303.5/PR308.2, and DBL5471/Fuvo. For details, please consult our technical team.